Standard Practice for Characterization and Verification of Linear Phased Array Ultrasonic Probes
Standard Practice for Characterization and Verification of Linear Phased Array Ultrasonic Probes
SIGNIFICANCE AND USE 5.1 This practice is intended to provide standardized procedures for evaluating linear phased-array ultrasonic probes. It is not intended to define performance and acceptance criteria, but rather to provide data from which such criteria may be established. 5.2 Implementation may require more detailed procedural instructions in a format of the using facility. 5.3 The measurement data obtained may be employed by users of this guide to specify, describe, or provide performance criteria for procurement and quality assurance, or service evaluation of the operating characteristics of linear phased-array ultrasonic probes. All or portions of the standard practice may be used as determined by the user. 5.4 The measurements are made primarily under pulse-echo conditions. To determine the relative performance of a probe element as either a transmitter or a receiver may require additional tests. 5.5 While these procedures relate to many of the significant parameters, others that may be important in specific applications may not be treated. These might include power handling capability, breakdown voltage, wear properties of contact units, radio-frequency interference, and the like. 5.6 Care must be taken to ensure that comparable measurements are made and that users of the standard practice follow similar procedures. The conditions specified or selected (if optional) may affect the test results and lead to apparent differences. 5.7 Interpretation of some test results, such as the shape of the frequency response curve, may be subjective. Small irregularities may be significant. Interpretation of the test results is beyond the scope of this practice. 5.8 Certain results obtained using the procedures outlined may differ from measurements made with phased-array ultrasonic test instruments. These differences may be attributed to differences in the nature of the experiment or the electrical characteristics of the instrumentation. 5.9 The pulse generator used t... SCOPE 1.1 This practice covers measurement procedures for evaluating certain characteristics of phased-array ultrasonic probes that are used with phased-array ultrasonic examination instrumentation. 1.2 This practice describes means for obtaining performance data that may be used to define the acoustic and electric responses of phased-array ultrasonic probes including contact (with or without a wedge) and immersion linear phased-array probes used for ultrasonic nondestructive testing with central frequencies ranging from 0.5 MHz to 10 MHz. Frequencies outside of this range may use the same methods but the testing equipment may vary. 1.3 When ultrasonic values dependent on material are specified in this practice, they are based on carbon steel with an ultrasonic wave propagation speed of 5920 m/s (±50 m/s) for longitudinal wave modes and 3255 m/s (±30 m/s) for transverse or shear wave modes. 1.4 This practice describes some of the characterization and verification procedures that can be carried out at the end stage of the manufacturing process of linear phased array probes. This practice does not describe the methods or acceptance criteria used to verify the performance of the combined phased array ultrasonic instrument and probe system. 1.5 While this practice is intended to provide standardized procedures for evaluating linear phased-array ultrasonic probes, it may, with suitable modifications, be used for evaluation of configurations other than linear; for example, 1.5D or 2D matrix array probes. 1.6 Units—The values stated in SI units are to be regarded as the standard. The values given in parentheses after SI units are provided for information only and are not considered standard. 1.7 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practice...
ASTM E2904-22: Standard Practice for Characterization and Verification of Linear Phased Array Ultrasonic Probes provides standardized measurement procedures for evaluating key physical, acoustic, and electrical characteristics of linear phased array ultrasonic probes. Developed by ASTM International’s Committee E07 on Nondestructive Testing, this standard ensures consistency and reliability in the assessment of phased array ultrasonic technology used with examination instrumentation. ASTM E2904 is designed to generate quantitative data that can help define performance specifications, ensure quality assurance, and facilitate the procurement and servicing of phased array probes, particularly in ultrasonic non-destructive testing (NDT) applications.
ASTM E2904-22 is widely applicable in various industries where ultrasonic non-destructive testing (NDT) is critical, such as:
Другие стандарты ASTM
The standard is not intended to establish acceptance or rejection criteria, but rather to inform and support the development of such criteria based on reliable, comparable data. Test conditions, including material types (commonly carbon steel), are standardized where possible to improve data consistency.
For enhanced utility and context, ASTM E2904-22 references and aligns with several related standards in ultrasonic nondestructive testing, including:
ASTM E2904-22 complements and builds on these documents by focusing specifically on phased array probe characterization, and by providing procedures relevant for a wide range of industries that utilize ultrasonic phased-array technology.
Keywords: ASTM E2904, phased array ultrasonic probes, ultrasonic nondestructive testing, probe characterization, probe verification, quality assurance, linear phased-array, NDT standards, acoustic measurement, probe performance.