ASTM E579-04(2023) PDF
Standard Test Method for Limit of Detection of Fluorescence of Quinine Sulfate in Solution
Standard Test Method for Limit of Detection of Fluorescence of Quinine Sulfate in Solution
- Статус документа:
- Действующий
- Формат:
- Электронный (PDF)
- Количество страниц:
- 3
- Дата публикации:
- 15 января 2023 г.
- Издание:
- E579
- ICS:
- 71.080.30
SIGNIFICANCE AND USE 4.1 When determining the limiting detectable concentration of a fluorescent substance, it is usually necessary to increase the readout scale of a photoelectric instrument to a point where noise (that is, random fluctuations of the system) becomes apparent. This noise will be superimposed upon the signal from the sample. 4.2 In molecular fluorescence spectroscopy, the limit of detection for the sample will be determined by the limiting signal-to-noise ratio, S/N, where the signal, S, is the difference between readings obtained with the sample and blank solutions, and N is the total root-mean-square (rms) noise. The limit of detection for the sample will be given by the instrument readings that give a signal equal to three times the rms value of the noise. Note 2: Factors other than noise affecting the sample concentration corresponding to the limit of detection include: the spectral bandwidths of the excitation and emission monochromators, the intensity of the exciting light that can be concentrated on the sample, the fraction of the fluorescence collected by the detection system, the response time of the detection system, and the purity of the solvent. The size and arrangement of the sample container with respect to the light beams are also important, as they affect both the desired signal and the extraneous signal that only contributes noise. Note 3: The value of rms noise (N) can be obtained by calculating the standard deviation of a series of readings of the signal from the sample at the peak emission wavelength at approximately 450 nm as follows: where: = mean of the series of readings, x = value of the individual reading, and n = number of readings. Alternatively, rms noise may be estimated by noting the extreme differences between the members of a series of readings (peak-to-peak noise) and dividing by a factor that is usually taken to be 5.6, 7 SCOPE 1.1 This test method employs the signal-to-noise ratio to determine the sensitivity of a fluorescence measuring system in testing for the limit of detection (LOD) of quinine sulfate dihydrate in solution. The results obtained with quinine sulfate dihydrate in solution are suitable for specifying instrument performance on samples having excitation and fluorescence bands wider than 10 nm at or near room temperature. 1.1.1 This test method is not intended to be used as (1) a rigorous test of performance of instrumentation, or (2), to intercompare the quantitative performance of instruments of different design. Intercomparison of the LOD between instruments is commonly expressed as the ratio of the water Raman peak intensity to the root-mean-square (rms) noise as measured on a fluorometer using an excitation wavelength of 350 nm This test method uses the excitation and emission peak wavelengths for quinine sulfate dihydrate in solution, which are approximately 350 nm and 450 nm, respectively. 1.2 This test method has been applied to fluorescence-measuring systems utilizing non-laser, low-energy excitation sources. There is no assurance that extremely intense illumination will not cause photodecomposition2 of the compound suggested in this test method. For this reason, it is recommended that this test method not be indiscriminately employed with high intensity light sources. This test method is not intended to determine minimum detectable amounts of other materials. If this test method is extended to employ other chemical substances, the user should be aware of the possibility that these other substances may undergo decomposition or adsorption onto containers. 1.3 A typical LOD for conventional fluorometers using this test method is 1 ng of quinine sulfate per mL. 1.4 The suggested shelf life of a 1 mg/mL stock solution of quinine sulfate dihydrate is three months, when stored in the dark in a stoppered glass bottle. 1.5 The values stated in SI units are to be regarded as standard. No o...
Abstract
Overview
ASTM E579-04(2023) is an international standard developed by ASTM International, titled Standard Test Method for Limit of Detection of Fluorescence of Quinine Sulfate in Solution. This standard provides a validated procedure for determining the limit of detection (LOD) of quinine sulfate dihydrate in solution using fluorescence-measuring systems. It focuses on using the signal-to-noise ratio (S/N) as a measure of sensitivity for molecular fluorescence spectroscopy and is suitable for evaluating instrument performance where excitation and fluorescence bandwidths exceed 10 nm at or near room temperature.
Key Topics
- Signal-to-Noise Ratio (S/N): The core concept is determining the LOD as the concentration at which the fluorescence signal is three times the root-mean-square (rms) noise of the system.
- Instrument Sensitivity: The method is intended to specify the sensitivity of fluorescence spectrometers, particularly regarding their ability to detect low concentrations of quinine sulfate.
- Operating Parameters: Critical factors affecting LOD include spectral bandwidths, light intensity, fraction of fluorescence collected, detector response time, solvent purity, and the configuration of the sample container relative to the light beams.
- Noise Calculation: The rms noise should be calculated from repeated readings at the emission peak (~450 nm) or estimated from peak-to-peak fluctuations.
- Solution Preparation: Involves preparing a stock solution of quinine sulfate dihydrate (1 mg/mL) and performing serial dilutions down to the ng/mL range.
Applications
ASTM E579-04(2023) is widely used in laboratories and industries that require accurate and reliable calibration of fluorescence spectrophotometers and fluorometers. Its primary applications include:
- Quality Control and Calibration: Ensuring fluorescence measuring equipment meets sensitivity requirements, especially when analyzing substances similar to quinine sulfate.
- Instrument Benchmarking: Characterizing instrument performance for samples at low concentration levels, without serving as a definitive comparison between instruments of differing design.
- Research and Development: Supporting the development of fluorescence techniques and photometric instrumentation in chemical, pharmaceutical, and environmental labs.
- Sample Analysis: Although developed for quinine sulfate, it provides guidance for adapting procedures for other compounds, with attention to possible solution decomposition or adsorption.
- Shelf Life Guidance: The standard includes recommendations for the shelf life and storage conditions of quinine sulfate stock solutions, supporting consistent results over time.
Related Standards
Organizations can refer to the following documents and standards for complementary guidelines or expanded methodologies:
- ASTM E578: Test Method for Linearity of Fluorescence Measuring Systems – cited for establishing operating wavelengths and instrument calibration procedures.
- Molecular Spectroscopy Standards: Further ASTM and ISO standards on UV-visible and luminescence spectroscopy support consistent methodology across analytical applications.
- WTO TBT Principles: ASTM E579-04(2023) aligns with international principles for the development of standards recognized by the World Trade Organization to facilitate trade and technical harmonization.
Practical Value
Implementing ASTM E579-04(2023) ensures that laboratories:
- Maintain a consistent and internationally recognized process for determining the fluorescence limit of detection.
- Achieve reliable, reproducible instrument calibration and performance verification.
- Navigate factors impacting LOD, including instrument configuration and environmental conditions.
- Remain compliant with best practices for molecular fluorescence spectroscopy in scientific and industrial settings.
Keywords: fluorescence limit of detection, quinine sulfate, spectroscopy, fluorometer calibration, signal-to-noise ratio, molecular spectroscopy standards, instrument sensitivity, ASTM E579
Технические детали
- Технический комитет
- E13 - Molecular Spectroscopy and Separation Science
- SKU
- ASTM E579-04(2023)
Похожие стандарты
Другие стандарты ASTM
ASTM-TPT-47
Phase I & Phase II Environmental Site Assessment Processes
Phase I & Phase II Environmental Site Assessment Processes
ASTM-TPT-8
Phase I Environmental Site Assessment Practices For Commercial Real Estate: Phase I Site Assessment & Transac…
ASTM-TPT-1137
ASTM D975 Standard Specification for Diesel Fuel
ASTM D975 Standard Specification for Diesel Fuel
ASTM-TPT-1202
ASTM D8421 Standard Test Method for Determination of Per- and Polyfluoroalkyl Substances (PFAS) in Aqueous Ma…
ASTM D8421 Standard Test Method -- eLearning Course
ASTM-TPT-691
Microaprendizagem para D4057: Amostragem de ponto (PT)
D4057 Microlearning: Spot Sampling Portuguese
ASTM-TPT-1091
Metodo de prueba estandar D4052 de ASTM -- Curso de aprendizaje electrónico (ES)
Modulo eLearning para ASTM D4052 en espanol
ASTM ACEM20180086
Using Neutron Radiography to Quantify the Settlement of Fresh Concrete
Specifications have been implemented for concrete bridge decks in North America that restrict the use of higher slump concrete mixtures primarily because of concerns about differential settlement and…
ASTM ACEM20180041
Experimental Study of Competing Failure of Reinforced Concrete Based on the Weibull Distribution
To predict the failure lifetime of reinforced concrete, a current accelerating corrosion test was performed on concrete by simulating the environment in an area with saline soil. To study the concret…