BS EN 60976:2007 PDF
Medical electrical equipment. Medical electron accelerators. Functional performance characteristics
Medical electrical equipment. Medical electron accelerators. Functional performance characteristics
- Статус документа:
- Действующий
- Формат:
- Электронный (PDF)
- Дата публикации:
- 30 марта 2004 г.
- ICS:
- 11.040.50
- Технический комитет:
- CENELEC
- SKU:
- BS EN 60976:2007
Abstract
What is BS EN 60976- P erformance characteristics of the medical electron accelerator about?
BS EN 60976 is a European standard on medical electrical equipment that discusses the performance characteristics of the medical electron accelerator to enhance the performance of the medical electron accelerators. BS EN 60976 applies to medical electron accelerators when used, for therapy purposes, in human medical practice. BS EN 60976 applies to medical electron accelerators which deliver a radiation beam of either X-radiation or Electron radiation with nominal energies in the range of 1 MeV to 50 MeV at maximum absorbed dose rates between 0,001 Gy s–1 and 1 Gy s–1 at 1 m from the Radiation source and at normal treatment distances between 50 cm and 200 cm from the radiation source. BS EN 60976 describes measurements and test procedures to be performed by the MANUFACTURER at the design and construction stage of a medical electron accelerators but does not specify acceptance test to be performed after installation at the purchaser's site. T...
Похожие стандарты