ASTM ISO/ASTM51956-21
Standard Practice for Use of a Thermoluminescence-Dosimetry System (TLD System) for Radiation Processing
Standard Practice for Use of a Thermoluminescence-Dosimetry System (TLD System) for Radiation Processing
- Статус документа:
- Действующий
- Формат:
- Электронный (PDF)
- Количество страниц:
- 8
- Дата публикации:
- 1 октября 2021 г.
- Издание:
- ISO/ASTM51956
- ICS:
- 17.240
SIGNIFICANCE AND USE 4.1 In radiation processing, TLDs are mainly used in the irradiation of blood products (see ISO/ASTM Practice 51939) and insects for sterile insect release programs (see ISO/ASTM Guide 51940). TLDs may also be used in other radiation processing applications such as the sterilization of medical products, food irradiation, modification of polymers, irradiation of electronic devices, and curing of inks, coatings and adhesives. (See ISO/ASTM Practices 51608, 51649, and 51702.) 4.2 For radiation processing, the absorbed-dose range of interest is from 1 Gy to 100 kGy. Some TLDs can be used in applications requiring much lower absorbed doses (for example, for personnel dosimetry), but such applications are outside the scope of this practice. Examples of TLDs and applicable dose ranges are given in Table 1. Information on various types of TLDs and their applications can be found in Refs (1-10).7 (A) This table is taken from Ref (6). Ranges are approximate, and may vary with batch. Supralinearity refers to a region where the slope of the response versus dose curve is greater than that for the linear region. 4.3 Information on other dosimetry systems used for radiation processing can be found in ICRU Report 80. SCOPE 1.1 This practice covers procedures for the use of thermoluminescence dosimeters (TLDs) to measure the absorbed dose in materials irradiated by photons or electrons in terms of absorbed dose to water. Thermoluminescence-dosimetry systems (TLD systems) are generally used as routine dosimetry systems. 1.2 The thermoluminescence dosimeter (TLD) is classified as a type II dosimeter on the basis of the complex effect of influence quantities on the dosimeter response. See ISO/ASTM Practice 52628. 1.3 This document is one of a set of standards that provides recommendations for properly implementing dosimetry in radiation processing, and describes a means of achieving compliance with the requirements of ISO/ASTM 52628 “Practice for Dosimetry in Radiation Processing” for a TLD system. It is intended to be read in conjunction with ISO/ASTM 52628. 1.4 This practice covers the use of TLD systems under the following conditions: 1.4.1 The absorbed-dose range is from 1 Gy to 10 kGy. 1.4.2 The absorbed-dose rate is between 1 × 10-2 and 1 × 1010 Gy s-1. 1.4.3 The radiation-energy range for photons and electrons is from 0.1 to 50 MeV. 1.5 This practice does not cover measurements of absorbed dose in materials subjected to neutron irradiation. 1.6 This practice does not cover procedures for the use of TLDs for determining absorbed dose in radiation-hardness testing of electronic devices or for clinical dosimetry. Procedures for the use of TLDs for radiation-hardness testing are given in ASTM Practice E668. Procedures for use of TLDs in clinical dosimetry are given in ISO 28057. 1.7 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use. 1.8 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
Abstract
Overview
ASTM ISO/ASTM51956-21: Standard Practice for Use of a Thermoluminescence-Dosimetry System (TLD System) for Radiation Processing establishes guidelines for using TLD systems to accurately measure absorbed radiation dose in materials processed with ionizing photons or electrons. Developed by ASTM, this international standard provides essential procedures to ensure effective and consistent dosimetry practices in the context of industrial and scientific radiation processing.
Thermoluminescence dosimeters (TLDs) are widely used as routine dosimetry systems due to their reliability and the ability to measure doses in a broad range-from 1 Gray (Gy) to 10 kilogray (kGy)-making them suitable for many crucial radiation processing applications. The standard outlines procedures for calibration, verification, use, and maintenance of TLDs to maintain measurement quality and meet regulatory and quality management requirements.
Key Topics
- Scope of Use: Specifies the use of TLD systems for absorbed doses between 1 Gy and 10 kGy, excluding clinical dosimetry, neutron irradiation, and hardness testing for electronics.
- TLD System Components: Covers the components-TLDs, TLD readers, calibration protocols, quality assurance, and routine procedures.
- Influence Quantities: Discusses factors such as temperature, humidity, light, and dose rate that can affect TLD system response and emphasizes the importance of controlling and documenting these variables.
- Calibration and Verification: Details on the importance of system calibration to process conditions, management of batch calibration, and the need for periodic verification to ensure ongoing accuracy.
- Measurement Management: Addresses preparation, storage, and handling of TLDs, plus the requirements for regular performance verification, documentation, and uncertainty evaluation.
- Uncertainty and Quality: Institutes the use of uncertainty budgets to identify and quantify all measurement uncertainties impacting the dosimetry results.
Applications
Thermoluminescence dosimetry as specified in ISO/ASTM51956-21 is integral to many radiation processing operations, including:
- Blood Product Irradiation: Ensures proper dosing for transfusion safety, in compliance with blood irradiation dosimetry standards.
- Sterile Insect Release Programs: Enables effective irradiation of insects for population control.
- Medical Product Sterilization: Assures sterilization of medical devices and pharmaceuticals through precise dose delivery.
- Food Irradiation: Provides verification of absorbed doses to ensure food safety and compliance with regulatory standards.
- Polymer Modification: Facilitates the control of dose delivery in processes that modify the properties of plastics and polymers.
- Curing of Inks, Coatings, and Adhesives: Supports the consistent treatment of materials where radiation-induced curing is required.
The standard also highlights the necessity of proper documentation, storage, and instrument maintenance, as well as the need to estimate and manage measurement uncertainties within these applications.
Related Standards
For comprehensive compliance and best practices, users of ISO/ASTM51956-21 should consult the following related standards:
- ISO/ASTM 52628: Dosimetry in Radiation Processing (companion standard covering general dosimetry system requirements)
- ISO/ASTM 51261: Calibration of Routine Dosimetry Systems for Radiation Processing
- ISO/ASTM 51608, 51649, 51702: Dosimetry in X-Ray, Electron-Beam, and Gamma Irradiation Facilities
- ISO/ASTM 51707: Uncertainty Estimation in Dosimetry for Radiation Processing
- ISO/ASTM 51939: Blood Irradiation Dosimetry
- ISO/ASTM 51940: Dosimetry for Sterile Insect Release
- ISO/ASTM 52701: Guide for Performance Characterization of Dosimeters and Systems
- ICRU Report 80: Dosimetry Systems for Use in Radiation Processing
Practical Value
Adhering to ASTM ISO/ASTM51956-21 ensures that radiation processing facilities can perform precise, reliable, and documented absorbed dose measurements, supporting regulatory compliance and high-quality assurance across industries such as healthcare, food safety, materials processing, and environmental management. Routine use of TLD systems, along with robust calibration, environmental controls, and systematic documentation, helps organizations minimize error, maintain safety, and achieve consistent output in all radiation processing applications.
Keywords: Thermoluminescence dosimeter, TLD system, radiation processing, absorbed dose measurement, calibration, quality assurance, industrial dosimetry, ionizing radiation, ASTM standards, routine dosimetry system.
Технические детали
- Технический комитет
- E61 - Radiation Processing
- SKU
- ASTM ISO/ASTM51956-21
Похожие стандарты
Другие стандарты ASTM
ASTM E7-22
ДействующийStandard Terminology Relating to Metallography
Overview ASTM E7-22 - Standard Terminology Relating to Metallography is a key terminology standard developed by ASTM Committee E04. It establishes uniform definitions for terms, acronyms, and symbols…
ASTM D56-22
ДействующийStandard Test Method for Flash Point by Tag Closed Cup Tester
Overview ASTM D56-22, Standard Test Method for Flash Point by Tag Closed Cup Tester, is a widely recognized method for determining the flash point of volatile liquids under controlled laboratory cond…
ASTM E18-22
ДействующийStandard Test Methods for Rockwell Hardness of Metallic Materials
Overview ASTM E18-22: Standard Test Methods for Rockwell Hardness of Metallic Materials defines procedures and requirements for determining the Rockwell and Rockwell superficial hardness of metallic…
ASTM E29-22
ДействующийStandard Practice for Using Significant Digits in Test Data to Determine Conformance with Specifications
Overview ASTM E29-22: Standard Practice for Using Significant Digits in Test Data to Determine Conformance with Specifications provides essential guidelines for the use of significant digits when est…
ASTM E11-22
ДействующийStandard Specification for Woven Wire Test Sieve Cloth and Test Sieves
Overview ASTM E11-22: Standard Specification for Woven Wire Test Sieve Cloth and Test Sieves establishes the technical requirements for woven wire sieve cloth and the construction of test sieves used…
ASTM ISO/ASTM51607-22
ДействующийStandard Practice for Use of an Alanine-EPR Dosimetry System
Overview ASTM ISO/ASTM51607-22 provides internationally recognized guidance on the implementation and use of the alanine-EPR dosimetry system to measure absorbed dose during radiation processing. Thi…
ASTM ISO/ASTM51401-21
ДействующийStandard Practice for Use of a Dichromate Dosimetry System
Overview ASTM ISO/ASTM51401-21: Standard Practice for Use of a Dichromate Dosimetry System establishes best practices for the preparation, testing, and operation of dichromate dosimeters to measure t…
ASTM ISO/ASTM51650-21
ДействующийStandard Practice for Use of a Cellulose Triacetate Dosimetry System
Overview ASTM ISO/ASTM 51650-21: Standard Practice for Use of a Cellulose Triacetate Dosimetry System establishes procedures for employing cellulose triacetate (CTA) dosimetry systems to measure abso…