ISO 10993-14:2001
Biological evaluation of medical devices — Part 14: Identification and quantification of degradation products from ceramics
Biological evaluation of medical devices — Part 14: Identification and quantification of degradation products from ceramics
- Статус документа:
- Действующий
- Формат:
- Электронный (PDF)
- Количество страниц:
- 11
- Дата публикации:
- 22 ноября 2001 г.
- Издание:
- ISO IS 10993 edition 1 version 1
- ICS:
- 11.100.20
La présente partie de l'ISO 10993 spécifie deux méthodes permettant d'obtenir des solutions de produits de dégradation à partir des céramiques (dont les verres) dans un but de quantification. Elle donne également des lignes directrices pour l'analyse de ces solutions afin d'identifier les produits de dégradation. Etant donné la nature générale de la présente partie de l'ISO 10993, il convient de prendre en compte en premier lieu les normes spécifiques aux produits qui, lorsqu'elles existent, abordent la formation de produits de dégradation dans des conditions d'utilisation plus pertinentes La présente partie de l'ISO 10993 aborde seulement les produits de dégradation générés par dissociation chimique des céramiques lors d'essais in vitro. Aucune dégradation due à une contrainte mécanique ou une énergie externe ne relève de la présente partie de l'ISO 10993. Alors que l'ISO 6872 et l'ISO 9693 sont des essais de dégradation chimique, elles ne traitent pas de l'analyse des produits de dégradation. Étant donné la gamme des céramiques utilisées dans les dispositifs médicaux et les différentes exigences d'exactitude et de fidélité des résultats, aucune technique d'analyse spécifique n'est identifiée. De plus, la présente partie de l'ISO 10993 ne fait état d'aucune exigence spécifique relative à des niveaux acceptables de produits de dégradation. Bien que ces matériaux soient conçus pour des applications biomédicales, l'activité biologique de ces produits de dégradation ne fait pas l'objet de la présente partie de l'ISO 10993.
Abstract
Overview
ISO 10993-14:2001 is part of the ISO 10993 series on the biological evaluation of medical devices. It specifies in vitro methods to obtain, identify and quantify degradation products from ceramics (including glasses and ceramic coatings). The standard defines two complementary test approaches - an extreme solution test (worst-case, low pH) and a simulation solution test (near in vivo pH) - and provides guidance on specimen preparation, testing conditions and analysis of the resulting solutions (filtrate) for risk assessment and further biological evaluation.
Key topics and technical requirements
- Test scope: Covers degradation by chemical dissociation in aqueous in vitro environments only; mechanical or enzyme/protein‑mediated degradation is excluded.
- Two test methods:
- Extreme solution test: Uses a citric acid buffer at pH 3.0 ± 0.2 (37 ± 1 °C) as a worst‑case environment; intended as a screening method for most ceramics.
- Simulation solution test: Designed to simulate more common in vivo pH conditions (details and decision flow outlined in the standard).
- Specimen preparation:
- Bulk, granular and coatings are covered; coatings may be removed and tested separately.
- Granulation by tungsten carbide mortar and pestle; target particle sizing to pass a 400 µm screen and be retained on 315 µm.
- Mass loading guidance: typically 5.00 ± 0.05 g for low‑solubility material and 10.00 ± 0.05 g for high‑solubility material (or scaled tests with justification).
- Test controls and equipment:
- Use 250 ml polypropylene or high‑density polyethylene containers (avoid glass).
- Filtration using a Büchner funnel to separate filtrate and retentate.
- Mass determination with balances accurate to 0.0005 g and six replicates; drying at 100 ± 2 °C until mass change ≤ 0.1%.
- Analysis of filtrate: Guidance on selecting which chemicals/elements to analyze and on required analytical sensitivity, but no prescriptive analytical techniques or acceptance limits - laboratories choose validated methods appropriate to required detection limits.
Applications and users
- Intended for medical device manufacturers, materials scientists, analytical testing laboratories and regulatory bodies assessing the chemical safety of ceramic implants, prosthetics, dental implants (note: dental ceramics exposed to the oral cavity should follow ISO 6872 for extreme tests), and ceramic coatings.
- Used to identify and quantify leachable ions and degradation products that form the basis for toxicological risk assessment and any subsequent biological safety testing under ISO 10993-1.
Related standards
- ISO 10993 series (notably ISO 10993-1, ISO 10993-9, ISO 10993-13, ISO 10993-15)
- ISO 6872 (dental ceramics) and reference standards for sieves, water and density (ISO 3310-1, ISO 3696, ISO 5017, ISO 6474)