ISO 14879-1:2020
Implants for surgery — Total knee-joint prostheses — Part 1: Determination of endurance properties of knee tibial trays
Implants for surgery — Total knee-joint prostheses — Part 1: Determination of endurance properties of knee tibial trays
- Статус документа:
- Отменён
- Формат:
- Электронный (PDF)
- Количество страниц:
- 9
- Дата публикации:
- 20 июля 2020 г.
- Издание:
- ISO IS 14879 edition 2 version 1
- ICS:
- 11.040.40
This document specifies a test method for determining the endurance properties, under specified laboratory conditions, of tibial trays used in knee-joint prostheses to support and secure the plastic articulating surface. It applies to tibial trays which cover both the medial and lateral plateaux of the tibia. The test method does not apply to tibial components manufactured solely from plastic materials. This document does not cover methods of examining and reporting the final condition of the test specimen; these can be the subject of agreement between the test laboratory and the parties submitting the specimen for test. NOTE Correlation of test results with in vivo performance has not been established.
Abstract
Overview
ISO 14879-1:2020 specifies a laboratory test method to determine the endurance properties of knee tibial trays used in total knee-joint prostheses. The standard defines how to fix a tibial tray as a cantilever, apply a cyclic compressive load through a spacer and spherical indentor, and record failure modes or runout. It applies to metal or metal-backed tibial trays that cover both medial and lateral tibial plateaux and explicitly excludes tibial components manufactured solely from plastic. ISO 14879-1:2020 notes that correlation with in vivo performance has not been established.
Key topics and technical requirements
- Test principle: Cantilever loading of one condyle with a cyclic force transmitted via a spacer; continue until fracture, cracking, excessive deflection, spacer failure, runout, or machine failure.
- Load control: Force-controlled testing with minimum force = 10% of maximum force; sinusoidal waveform recommended.
- Test frequency: Up to 10 Hz by default; higher frequencies may be used only with validated dynamic calibration showing equivalent results.
- Spacer and indenter:
- Recommended spacer diameter 13.0 mm ± 0.5 mm, minimum thickness 6.0 mm ± 0.5 mm; replace if thickness reduces to 3 mm or it fragments.
- Spherical load applicator with 16 mm radius recommended; spacer may be glued to prevent translation.
- Fixture and positioning:
- Support up to the anteroposterior centreline; maintain linear displacement tolerance ±1 mm and angular tolerance ±2°.
- Load axis perpendicular to the undeflected superior surface within 2°.
- Determine loading point and measure ML moment arm (d_ml) and AP moment arm (d_ap) (definitions per ISO 7207-1).
- Instrumentation and reporting:
- Force accuracy within ±2% at maximum (per ISO 7500-1).
- Record cycles to failure/runout, mode/location of failure, deflection, spacer use, test environment, and specimen ID.
Practical applications and users
- Product design and validation for knee implants (manufacturers and R&D teams)
- Independent mechanical testing laboratories performing fatigue and endurance testing
- Regulatory submissions and conformity evidence for total knee-joint prostheses
- Quality assurance and comparative evaluation of tibial tray designs and materials
- Failure analysis and pre-clinical performance benchmarking
Keywords: ISO 14879-1, knee tibial trays, endurance properties, tibial tray testing, total knee-joint prostheses, fatigue testing, spacer, load applicator, ML moment arm, AP moment arm.
Related standards
- ISO 7207-1 - classification, definitions and dimensions for knee joint components (referenced for centrelines and dimensions)
- ISO 7500-1 - calibration and verification of force-measuring systems (force accuracy)
- ISO 5833 - bone cement (embedding material reference)
- ISO 21536 - referenced endurance requirement (example runout guidance)
For implementation, consult the full ISO 14879-1:2020 text and coordinate test conditions and specimen reporting with the testing laboratory.
Технические детали
- Технический комитет
- ISO/TC 150/SC 4 - Bone and joint replacements
- SKU
- ISO 14879-1:2020
Похожие стандарты
Стандарты, упомянутые в описании
ISO 7207-1:2007
ДействующийImplants for surgery — Components for partial and total knee joint prostheses — Part 1: Classification, defin…
Overview ISO 7207-1:2007 - Implants for surgery - Components for partial and total knee joint prostheses - Part 1: Classification, definitions and designation of dimensions provides a standardized fr…
ISO 7500-1:2018
ДействующийMetallic materials — Calibration and verification of static uniaxial testing machines — Part 1: Tension/compr…
Overview ISO 7500-1:2018 - "Metallic materials - Calibration and verification of static uniaxial testing machines - Part 1: Tension/compression testing machines - Calibration and verification of the…
BS ISO 14879-1:2020
ДействующийImplants for surgery. Total kneejoint prostheses. Determination of endurance properties of knee tibial trays.
ISO 5833:2002
ДействующийImplants for surgery — Acrylic resin cements
Overview ISO 5833:2002 - "Implants for surgery - Acrylic resin cements" sets international requirements for acrylic resin (bone) cements used to fix internal orthopaedic prostheses. The standard appl…