ISO 15882:2008
Sterilization of health care products — Chemical indicators — Guidance for selection, use and interpretation of results
Sterilization of health care products — Chemical indicators — Guidance for selection, use and interpretation of results
- Статус документа:
- Действующий
- Формат:
- Электронный (PDF)
- Количество страниц:
- 31
- Дата публикации:
- 21 августа 2008 г.
- Издание:
- ISO IS 15882 edition 2 version 1
- ICS:
- 11.080.01
ISO 15882:2008 provides guidance for the selection, use and interpretation of results of chemical indicators used in process definition, validation and routine monitoring and overall control of sterilization processes. ISO 15882:2008 applies to indicators that show exposure to sterilization processes by means of physical and/or chemical change of substances, and which are used to monitor one or more of the variables required for a sterilization process. These chemical indicators are not dependent for their action on the presence or absence of a living organism.
Abstract
ISO 15882:2008 - Overview
ISO 15882:2008 gives practical guidance for the selection, use and interpretation of chemical indicators used to monitor sterilization of health care products. It applies to non‑biological indicators that change physically and/or chemically to show exposure to sterilization processes (steam, dry heat, ethylene oxide, γ/β radiation, low‑temperature steam/formaldehyde, vapourized hydrogen peroxide). The standard is intended for process definition, validation and routine monitoring but does not replace biological indicators or full sterilization validation requirements.
Key topics and requirements
- Scope & definitions: Clear definitions for terms such as chemical indicator, endpoint, stated value (SV), process variable/parameter, process challenge device (PCD) and process challenge location (PCL).
- Classes of chemical indicators: Guidance on Class 1–6 indicators (process indicators, class‑specific tests, single‑variable, multi‑variable, integrating and emulating indicators) and how class selection affects the information provided.
- Selection criteria: Match indicator class and stated value to the sterilization method, critical variables and the intended monitoring purpose.
- Use and placement: Emphasizes representative placement of indicators in loads and PCDs; selection of locations should be informed by validation studies.
- Interpretation of results: How to interpret visible change, graduated or endpoint responses and actions to take on indicator “fail” outcomes (investigate sterilizer, load configuration, maintenance, utilities, handling, etc.).
- Integration with sterility assurance: Role of chemical indicators within broader sterility assurance and their relationship to biological indicators and ISO performance standards.
- Operational controls: Recommendations on record keeping, personnel training, storage/handling and labelling/marking of indicators and packages.
- Informative annexes: Background on specific tests (e.g., Bowie‑Dick), terms explanation and rationale linking chemical indicators to biological indicator performance.
Practical applications and users
ISO 15882 is used by:
- Sterilization technicians and central sterile services departments (CSSD)
- Medical device manufacturers and packaging engineers
- Quality assurance and regulatory compliance teams
- Sterilizer manufacturers, validation specialists and service providers
Typical applications:
- Routine cycle monitoring and load release
- Process definition and validation studies
- Placement and use of indicators in process challenge devices (PCDs)
- Supporting investigations when chemical indicators show failed or unexpected responses
Related standards
- ISO 11140 series (chemical indicator performance and definitions, incl. ISO 11140‑1)
- ISO 14937 (general sterilization validation)
- ISO 17665 series (moist heat sterilization)
- ISO 11135 series (ethylene oxide sterilization)
- ISO 11137‑1 (radiation sterilization)
- ISO 20857 (dry heat sterilization)
- ISO/TS 11139 and ISO 11138 series (vocabulary, biological indicators)
Keywords: ISO 15882, chemical indicators, sterilization monitoring, sterilization validation, healthcare sterilization, process challenge device, PCD, sterility assurance.
Технические детали
- Технический комитет
- ISO/TC 198 - Sterilization of health care products
- SKU
- ISO 15882:2008
Похожие стандарты
Стандарты, упомянутые в описании
BS EN ISO 15882:2008
ДействующийSterilization of health care products. Chemical indicators. Guidance for selection, use and interpretation of…
BS EN ISO 11140-4:2007
ДействующийSterilization of health care products. Chemical indicators. Class 2 indicators as an alternative to the Bowie…
BS EN ISO 11140-1:2014
ДействующийSterilization of health care products. Chemical indicators. General requirements.
BS EN ISO 14937:2009
ДействующийSterilization of health care products. General requirements for characterization of a sterilizing agent and t…
ISO 17665:2024
ДействующийSterilization of health care products — Moist heat — Requirements for the development, validation and routine…
Overview - ISO 17665:2024 (Moist heat sterilization) ISO 17665:2024, "Sterilization of health care products - Moist heat - Requirements for the development, validation and routine control of a steril…
ISO 11135:2014
ДействующийSterilization of health-care products — Ethylene oxide — Requirements for the development, validation and rou…
Overview ISO 11135:2014 - Sterilization of health‑care products - Ethylene oxide specifies the requirements for the development, validation and routine control of an ethylene oxide (EO) sterilization…