ISO 16900-12:2016
Respiratory protective devices — Methods of test and test equipment — Part 12: Determination of volume-averaged work of breathing and peak respiratory pressures
Respiratory protective devices — Methods of test and test equipment — Part 12: Determination of volume-averaged work of breathing and peak respiratory pressures
- Статус документа:
- Действующий
- Формат:
- Электронный (PDF)
- Количество страниц:
- 15
- Дата публикации:
- 20 мая 2016 г.
- Издание:
- ISO IS 16900 edition 1 version 1
- ICS:
- 13.340.30
ISO 16900-12:2016 specifies the test methods for determining the volume-averaged work of breathing and peak respiratory pressures imposed by the respiratory protective device (RPD). Elastic work, elastic physiological effects, and information on physiological effects of work of breathing (WOB) are specified in ISO 16976‑4 and are not included in this test method.
Abstract
Overview
ISO 16900-12:2016 specifies standardized test methods for determining the volume-averaged work of breathing (WOB/V) and peak respiratory pressures imposed by a respiratory protective device (RPD). The standard defines the principle, required apparatus, data acquisition rates, validation steps and calculation methods used to quantify pressure–volume behavior of complete RPDs or respiratory interfaces. Note: elastic work and physiological effects of WOB are covered separately in ISO 16976‑4 and are not part of this test method.
Key topics and requirements
- Scope and purpose: Measurement of WOB (work of breathing) per breath and peak inspiratory/expiratory pressures for RPDs.
- Test apparatus: Use a breathing machine or metabolic simulator compliant with ISO 16900‑5, an appropriate RPD headform/torso, pressure transducers and a displacement sensor for piston volume measurement.
- Signal capture: Pressure and volume signals acquired at a minimum of 100 Hz per channel; pressure measurement system must handle frequencies up to 50 Hz with <3 dB damping.
- Validation: Verification using specified validation orifices (ISO 16900‑5) and the verification limits in the standard (Table 1) to ensure the breathing machine and setup produce acceptable WOB results.
- Test procedure: Fit RPD to headform, run at specified minute ventilation/tidal volume/frequency, ensemble-average 10 consecutive breaths after stabilization, then calculate WOB/V and peak pressures.
- Reporting & uncertainty: Test report requirements and application of measurement uncertainty are specified (Annex A). Ambient test conditions and tolerances (e.g., temperature range and ±5 % general tolerance) are defined.
Practical applications
- Product development and design validation for respirators, powered air-purifying respirators (PAPRs) and escape devices.
- Performance and conformity testing used by certification bodies and independent test laboratories to demonstrate compliance with respiratory protective device performance standards.
- Quality control in manufacturing to monitor breathing resistance and pressure performance over production runs.
- Research into respiratory ergonomics and comparative evaluation across RPD models.
Who should use this standard
- RPD manufacturers and design engineers
- Test laboratories and conformity assessment bodies
- Occupational safety professionals and procurement/specification teams
- Regulatory authorities and standards committees
Related standards
- ISO 16900 series (Parts 1–14) - broader RPD test methods
- ISO 16900‑5 - breathing machine, metabolic simulator, headforms and verification tools
- ISO 16976‑4 - physiological effects and elastic work (complementary to ISO 16900‑12)
Keywords: ISO 16900-12:2016, work of breathing, WOB, respiratory protective devices, peak respiratory pressures, breathing machine, RPD testing, validation orifices.
Технические детали
- Технический комитет
- ISO/TC 94/SC 15 - Respiratory protective devices
- SKU
- ISO 16900-12:2016
Похожие стандарты
Стандарты, упомянутые в описании
BS ISO 16976-4:2023
ДействующийRespiratory protective devices. Human factors. Work of breathing and breathing resistance: physiologically ba…
ISO 16900-5:2016
ДействующийRespiratory protective devices — Methods of test and test equipment — Part 5: Breathing machine, metabolic si…
Overview ISO 16900-5:2016 - Respiratory protective devices defines standardized test equipment and fixtures used in laboratory testing of respiratory protective devices (RPD). This part of ISO 16900…
ISO 16900-13:2015
ДействующийRespiratory protective devices — Methods of test and test equipment — Part 13: RPD using regenerated breathab…
Overview - ISO 16900-13:2015 (Respiratory protective devices) ISO 16900-13:2015 defines consolidated laboratory test methods for respiratory protective devices (RPD) that use regenerated breathable g…