ISO 17218:2014
Sterile acupuncture needles for single use
Sterile acupuncture needles for single use
- Статус документа:
- Действующий
- Формат:
- Электронный (PDF)
- Количество страниц:
- 24
- Дата публикации:
- 3 февраля 2014 г.
- Издание:
- ISO IS 17218 edition 1 version 1
- ICS:
- 11.040.25
ISO 17218:2014 specifies the requirements for the sterile acupuncture needles for single use (specialized for filiform needles).
Abstract
Overview
ISO 17218:2014 - Sterile acupuncture needles for single use specifies requirements for single‑use, manufacturer‑sterilized filiform acupuncture needles. The standard covers needle configuration, types (with or without guide tube), handle designs, materials, dimensional tolerances, performance, sterility assurance, packaging, labelling, transit/storage and test methods. It is intended to support safe, consistent manufacture and clinical use of sterile disposable acupuncture needles.
Key topics and technical requirements
- Scope and configuration
- Applies to filiform acupuncture needles for single use, sterilized by the manufacturer.
- Defines parts of the needle: tip, body, root, handle and tail; needle types with/without guide tube; common handle styles (ring, plain, flower, metal tube, plastic).
- Materials and biocompatibility
- Needle body biocompatibility must be assessed per ISO 10993‑1. Evidence may be based on published data, prior clinical use, traceability or specified biological tests (cytotoxicity, sensitization, intracutaneous reactivity, EO residuals).
- Common needle body materials noted (examples): X5CrNi18‑9, X7CrNi18‑9 stainless steels (see ISO 15510).
- Dimensions and tolerances
- Size designation is by nominal diameter × length (e.g., Ø 0.30 × 40 mm).
- Example diameter tolerances:
- 0.10 ≤ d < 0.25 mm: ±0.008 mm
- 0.25 ≤ d ≤ 0.45 mm: ±0.015 mm
- 0.45 < d ≤ 0.80 mm: ±0.020 mm
- Length tolerances vary with range (e.g., 5–25 mm ±0.50 mm; 25–60 mm ±1.00 mm; larger ranges with larger tolerances).
- Minimum handle length: 13 mm.
- Performance and sterility
- Surface cleanliness, straightness, absence of defects on body and handle.
- Sharpness and puncture performance are emphasized; Annex A (tip sharpness/intensity) and Annex B (puncture tests) provide evaluation methods. Annex C covers corrosion resistance testing.
- Sterility assurance requirements and package integrity for primary/secondary packaging are specified.
- Labelling, transit and storage
- Labelling requirements for both primary and secondary packages and guidance for transit/storage conditions to preserve sterility.
Practical applications - who uses this standard
- Medical device and needle manufacturers: product design, material selection, QC tolerances and test methods.
- Quality, regulatory and compliance teams: conformity evidence, sterility assurance, biocompatibility documentation.
- Testing laboratories: implement Annex A–C test methods for tip performance and corrosion.
- Procurement and clinicians (acupuncturists, clinics, hospitals): product specifications, safe single‑use selection and labeling interpretation.
- Notified bodies/regulators: reference for device evaluation and market surveillance.
Related standards
- ISO 10993‑1 - Biological evaluation of medical devices
- ISO 14971 - Risk management for medical devices (recommended framework)
- ISO 15223‑1 - Symbols for medical device labelling
- ISO 6507‑1 - Vickers hardness test (referenced)
- ISO 15510 - Materials (referenced for commonly used stainless steels)
Keywords: ISO 17218:2014, sterile acupuncture needles, single use, filiform needles, needle dimensions, sterility assurance, biocompatibility, puncture performance, acupuncture needle standards.
Технические детали
- Технический комитет
- ISO/TC 249 - Traditional Chinese medicine
- SKU
- ISO 17218:2014
Похожие стандарты
Стандарты, упомянутые в описании
ISO 10993-12:2012
ОтменёнBiological evaluation of medical devices — Part 12: Sample preparation and reference materials
BS ISO 15510:2014
ДействующийStainless steels. Chemical composition.
ISO 14971:2007
ДействующийMedical devices - Application of risk management to medical devices
Overview ISO 14971:2007 - Medical devices - Application of risk management to medical devices - defines a systematic process for identifying hazards, estimating and evaluating risks, implementing ris…