This document specifies test methods and procedures for assessing the accuracy of a three dimensional numerical description of intra-oral surfaces acquired directly from a patient with a hand-held scanning device. The test methods are not applicable to ultrasonic, radiographic or magnetic resonance imaging methods. NOTE ISO 12836 specifies the test methods for the assessment of accuracy of digitizing devices that use a fixed or a mechanically guided scanning device.
Abstract
Overview
ISO 20896-1:2019 - "Dentistry - Digital impression devices - Part 1: Methods for assessing accuracy" defines standardized test methods and procedures to evaluate the accuracy of three-dimensional digital impressions acquired directly from a patient using a hand-held intra‑oral scanning device. The standard covers test object preparation, reference measurement, scanning conditions and statistical treatment needed to determine trueness (bias) and precision (standard deviation) of intraoral scanners. It does not apply to ultrasonic, radiographic or MRI imaging.
Key topics and requirements
- Scope and definitions: Clarifies terms such as digital impression, hand‑held scanning device, range, registration, scanning pattern and dimension of interest.
- Test objects: Normative annexes specify representative test objects (e.g., crown preparation, inlay preparation, full‑arch) to reflect clinical scanning challenges.
- Reference measurement: Dimensions of interest must be measured independently with traceability to an international length standard; measurement uncertainty (standard uncertainty σ) must be recorded.
- Test conditions: Requirements for ambient conditions, prescribed scanning pattern(s), scanning time and number of scans to ensure repeatable evaluation of intraoral scanners.
- Data processing: Rules for processing digital impressions, defining dimensions of interest and how to determine them from the scan data.
- Assessment of accuracy: Statistical analysis to express bias (difference from reference) and accuracy (combination of trueness and precision); precision is operationally defined as standard deviation.
- Test report: Mandatory reporting elements include device identification, test object description, test method, environmental conditions, processing steps and numerical results.
Practical applications and users
ISO 20896-1 is essential for:
- Manufacturers of intraoral scanners for product development, validation and regulatory submissions.
- Dental laboratories and CAD/CAM centers validating scanner performance for restorative workflows (crowns, inlays, bridges, full‑arch prostheses).
- Clinical researchers and universities conducting comparative performance studies of intraoral scanners.
- Regulatory bodies and procurement teams needing objective, comparable accuracy data to evaluate devices.
- Test laboratories performing independent conformity and performance testing.
By standardizing test objects, scanning procedures and statistical reporting, ISO 20896-1 helps ensure reproducible, comparable assessments of digital impression device accuracy-a critical factor for reliable digital dental workflows and high‑quality restorations.
Related standards
- ISO 12836 (digitizing devices with fixed scanning)
- ISO 18739 (dentistry CAD/CAM vocabulary)
- ISO 5725‑1 (accuracy - trueness and precision)
- ISO/IEC Guide 98‑3 (GUM: uncertainty of measurement)
Технические детали
- Технический комитет
- ISO/TC 106/SC 9 - Dental CAD/CAM systems
- SKU
- ISO 20896-1:2019
Похожие стандарты
Стандарты, упомянутые в описании
BS EN ISO 20896-1:2019
ДействующийDentistry. Digital impression devices. Methods for assessing accuracy.
BS EN ISO 12836:2015
ДействующийDentistry. Digitizing devices for CAD/CAM systems for indirect dental restorations. Test methods for assessin…
BS EN ISO 18739:2026
ДействующийDentistry — Vocabulary of process chain for CAD/CAM systems.
ISO 5725-1:2023
ДействующийAccuracy (trueness and precision) of measurement methods and results — Part 1: General principles and definit…
Overview ISO 5725-1:2023, "Accuracy (trueness and precision) of measurement methods and results - Part 1: General principles and definitions," is published by the International Organization for Stand…