ISO 7206-10:2018
Implants for surgery — Partial and total hip-joint prostheses — Part 10: Determination of resistance to static load of modular femoral heads
Implants for surgery — Partial and total hip-joint prostheses — Part 10: Determination of resistance to static load of modular femoral heads
- Статус документа:
- Действующий
- Формат:
- Электронный (PDF)
- Количество страниц:
- 11
- Дата публикации:
- 9 августа 2018 г.
- Издание:
- ISO IS 7206 edition 2 version 1
- ICS:
- 11.040.40
This document specifies methods of determining the compressive (fracture) or the tension (disassembly) loads required, under specific laboratory conditions, to cause failure of a modular head system. This document applies to components made of metallic and non-metallic materials, such as femoral heads of partial or total hip-joint replacements of modular construction (i.e. a head/neck conical taper connection). This document excludes methods of examining and reporting the test specimens.
Abstract
Overview
ISO 7206-10:2018 specifies laboratory methods for determining the resistance to static load of modular femoral heads used in partial and total hip-joint prostheses. The standard defines how to measure the compressive (fracture) and tensile (disassembly) loads required to cause failure of a head/neck modular connection (conical or Morse taper) under controlled static conditions. It applies to components made of metallic and non‑metallic materials but excludes methods for examining and reporting test specimens.
Key topics and technical requirements
- Test objectives: Determine the axial compressive load that causes fracture and the axial tensile load that causes disassembly of the head/neck assembly.
- Test types: Static compression test and static tension test with specific apparatus and alignment requirements.
- Apparatus accuracy: Compression machines per ISO 7500-1 with force-measuring accuracy typically ±2% (compression) and ±1% (tension).
- Fixtures and loading:
- Conical loading bore (metal, hardness ≥ 150 HB) and a soft copper ring for load distribution and surface protection.
- Load application aligned with the neck axis; fixtures must avoid introducing bending moments or torque.
- Assembly and installation force:
- Clean components; hand-fit then apply an installation force of (2.0 ± 0.2) kN at specified rates before testing.
- Specimen requirements:
- Test heads must be actual implants; cones and bores must meet manufacturer’s tolerances for basic design controls and 3D control elements (engagement length, surface roughness, circularity, straightness, cone/bore angles, concentricity, etc.).
- Minimum sample size: at least five specimens of each type per static test.
- Procedure principles: Increase axial load until failure or a predefined maximum is reached; alignment and measurement procedures are defined in the standard’s figures and clauses.
Applications and who uses it
- Implant manufacturers: validate modular head/neck designs, production control and assembly procedures.
- Test laboratories and QA teams: perform conformity and batch testing for mechanical integrity of femoral heads.
- Regulatory bodies and notified bodies: assess mechanical safety evidence for pre-market clearance and post-market surveillance.
- Design engineers and R&D: compare materials (metallic vs non-metallic) and taper design choices, and optimize taper geometry and surface treatments.
- Clinical risk assessors: use results to inform clinical performance expectations and failure-mode analyses.
Keywords: ISO 7206-10, modular femoral heads, static load testing, hip-joint prostheses, compression test, tension test, Morse taper, implant testing.
Related standards
- ISO 7206-1 - Classification and designation of dimensions for hip prostheses
- ISO 7500-1 - Calibration and verification of static uniaxial testing machines
- ISO 6506-1 - Brinell hardness test (referenced for loading-bore material)
- ISO 4288 - Surface texture (profile method)
Технические детали
- Технический комитет
- ISO/TC 150/SC 4 - Bone and joint replacements
- SKU
- ISO 7206-10:2018
Похожие стандарты
Стандарты, упомянутые в описании
ISO 7500-1:2018
ДействующийMetallic materials — Calibration and verification of static uniaxial testing machines — Part 1: Tension/compr…
Overview ISO 7500-1:2018 - "Metallic materials - Calibration and verification of static uniaxial testing machines - Part 1: Tension/compression testing machines - Calibration and verification of the…
ISO 7206-1:1995
ОтменёнImplants for surgery — Partial and total hip joint prostheses — Part 1: Classification and designation of dim…
ISO 6506-1:2014
ДействующийMetallic materials — Brinell hardness test — Part 1: Test method
Overview ISO 6506-1:2014 - Metallic materials - Brinell hardness test - Part 1: Test method defines the standardized procedure for performing the Brinell hardness test on metallic materials using bot…