ISO 80601-2-74:2026
Medical electrical equipment — Part 2-74: Particular requirements for basic safety and essential performance of respiratory humidifying equipment
Medical electrical equipment — Part 2-74: Particular requirements for basic safety and essential performance of respiratory humidifying equipment
- Статус документа:
- Действующий
- Формат:
- Электронный (PDF)
- Количество страниц:
- 113
- Дата публикации:
- 10 апреля 2026 г.
- Издание:
- ISO IS 80601 edition 3 version 1
- ICS:
- 11.040.10
This document applies to the basic safety and essential performance of a humidifier, also hereafter referred to as ME equipment, in combination with its accessories, the combination also hereafter referred to as ME system. This document is also applicable to those accessories intended by their manufacturer to be connected to a humidifier where the characteristics of those accessories can affect the basic safety or essential performance of the humidifier. EXAMPLE 1 Heated breathing tubes (heated-wire breathing tubes) or ME equipment intended to control these heated breathing tubes (heated breathing tube controllers). NOTE 2 Heated breathing tubes and their controllers are ME equipment and are subject to the requirements of IEC 60601‑1. NOTE 3 ISO 5367 specifies other safety and performance requirements for breathing tubes. This document includes requirements for the different medical uses of humidification, such as invasive ventilation, non-invasive ventilation, nasal high-flow therapy, and obstructive sleep apnoea therapy, as well as humidification therapy for tracheostomy patients. NOTE 4 A humidifier can be integrated into other equipment. When this is the case, the requirements of the other equipment also apply to the humidifier. EXAMPLE 2 Heated humidifier incorporated into a critical care ventilator where ISO 80601‑2-12 also applies. EXAMPLE 3 Heated humidifier incorporated into a homecare ventilator for dependent patients where ISO 80601‑2‑72 also applies. EXAMPLE 4 Heated humidifier incorporated into sleep apnoea therapy equipment where ISO 80601‑2‑70 also applies. EXAMPLE 5 Heated humidifier incorporated into ventilatory support equipment where either ISO 80601-2-79 or ISO 80601-2-80 also apply. EXAMPLE 6 Heated humidifier incorporated into respiratory high-flow therapy equipment where ISO 80601‑2‑90 also applies. This document also includes requirements for an active HME (heat and moisture exchanger), ME equipment which actively adds heat and moisture to increase the humidity level of the gas delivered from the HME to the patient. This document is not applicable to a passive HME, which returns a portion of the expired moisture and heat of the patient to the respiratory tract during inspiration without adding heat or moisture. NOTE 5 ISO 9360‑1 and ISO 9360‑2 specify safety and performance requirements for a passive HME. NOTE 6 If a clause or subclause is specifically intended to be applicable to ME equipment only, or to ME systems only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to ME equipment and to ME systems, as relevant. Hazards inherent in the intended physiological function of ME equipment or ME systems within the scope of this document are not covered by specific requirements in this document except in IEC 60601‑1:2005+AMD1:2012+AMD2:2020, 7.2.13 and 8.4.1. NOTE 7 Additional information can be found in IEC 60601‑1:2005+AMD1:2012+AMD2:2020, 4.2. This document does not specify the requirements for cold pass-over or cold bubble-through humidification devices, the requirements for which are given in ISO 20789. This document is not applicable to equipment commonly referred to as “room humidifiers” or humidifiers used in heating, ventilation and air conditioning systems, or humidifiers incorporated into infant incubators to humidify the chamber air (i.e., are not directly connected to the patient). This document is not applicable to nebulizers used for the delivery of a drug to patients. NOTE 8 ISO 27427 specifies the safety and performance requirements for nebulizers.
Abstract
Overview
ISO 80601-2-74:2026 is an international standard developed by ISO that specifies the particular requirements for basic safety and essential performance of respiratory humidifying equipment in medical environments. This standard applies to electrical medical equipment (ME equipment) such as humidifiers and their related systems and accessories, especially where these can affect the safety or performance of the device. The requirements are designed to ensure the effective and safe use of humidification in respiratory care across both professional and home healthcare settings.
Key Topics
- Scope and Applicability: The standard covers humidifiers intended for use in combination with other respiratory care equipment, including accessories like heated breathing tubes and temperature sensors. It applies to devices used for invasive ventilation, non-invasive ventilation, nasal high-flow therapy, obstructive sleep apnoea therapy, and tracheostomy humidification.
- Active vs. Passive Devices: ISO 80601-2-74:2026 is specific to active humidifiers and active heat and moisture exchangers (HMEs) that add heat or moisture to respiratory gases. Passive HMEs and room humidifiers are excluded from the scope of this standard.
- Safety and Performance: There is a strong focus on minimizing electrical, mechanical, and radiation hazards. The standard details requirements for alarm systems, controls accuracy, temperature limitations, and system recovery protocols.
- System Integration: Humidifiers may be integrated into larger respiratory support systems (such as ventilators or sleep apnoea equipment). In these cases, additional standards may apply.
- Accessory Requirements: Any accessory or component that could affect the basic safety or essential function of a humidifier falls within the standard’s requirements.
- Labeling and Documentation: The standard outlines comprehensive requirements for marking, labelling, and accompanying information to ensure correct, safe usage by end-users.
- Testing and Classification: Detailed procedures for classification, testing, and validation are described to ensure continued safety and compliance of the equipment.
Applications
ISO 80601-2-74:2026 supports medical device manufacturers, regulatory authorities, and clinical users by providing a framework for the design, testing, and selection of respiratory humidification equipment. Practical applications include:
- Critical Care: Ensuring that heated humidifiers integrated in intensive care ventilators conform to stringent safety and performance criteria.
- Home Healthcare: Providing basic safety requirements for portable or home-use humidifiers to protect vulnerable patient populations.
- Respiratory Therapy: Guiding the selection and validation of equipment for nasal high-flow therapy, sleep apnoea treatment, and tracheostomy care.
- Compatibility Assessment: Assisting healthcare providers and manufacturers in choosing safe and compatible accessories, such as breathing tubes and controllers, that interact with the primary humidifier unit.
Incorporation of this standard into procurement, risk assessment, and clinical workflow ensures safer respiratory care and regulatory compliance.
Related Standards
To ensure comprehensive safety and compatibility, reference to additional standards is often necessary. Key related standards include:
- IEC 60601-1: General requirements for basic safety and essential performance of medical electrical equipment.
- ISO 5367: Safety and performance requirements for breathing tubes.
- ISO 20789: Requirements for cold pass-over and cold bubble-through humidifiers.
- ISO 9360-1 & ISO 9360-2: Requirements for passive heat and moisture exchangers.
- ISO 27427: Safety and performance requirements for nebulizers.
- IEC 60601-1-2, 60601-1-6, 60601-1-8, 60601-1-11: Collateral standards covering electromagnetic compatibility, usability, alarms, and home use, respectively.
For medical device compliance and effective implementation, referencing ISO 80601-2-74:2026 alongside these standards ensures safe, reliable humidification in respiratory care across all healthcare settings.
Технические детали
- Технический комитет
- ISO/TC 121/SC 3 - Respiratory devices and related equipment used for patient care
- SKU
- ISO 80601-2-74:2026
Похожие стандарты
Стандарты, упомянутые в описании
IEC TR 62348:2012
ДействующийAssessment of the impact of the most significant changes in Amendment 1 to IEC 60601-1:2005 and mapping of th…
Overview IEC TR 62348:2012 is a Technical Report from the International Electrotechnical Commission that helps users transition from IEC 60601-1:2005 (Edition 3) to the amended edition (Amendment 1:2…
ISO 5367:2014
ОтменёнAnaesthetic and respiratory equipment — Breathing sets and connectors
BS EN ISO 20789:2019
ДействующийAnaesthetic and respiratory equipment. Passive humidifiers.
ISO 9360-1:2000
ДействующийAnaesthetic and respiratory equipment — Heat and moisture exchangers (HMEs) for humidifying respired gases in…
Overview ISO 9360-1:2000 defines performance and test requirements for heat and moisture exchangers (HMEs) used to humidify respired gases in humans for patients with minimum tidal volumes of 250 ml.…
ISO 9360-2:2001
ДействующийAnaesthetic and respiratory equipment — Heat and moisture exchangers (HMEs) for humidifying respired gases in…
Overview ISO 9360-2:2001 specifies requirements and test methods for heat and moisture exchangers (HMEs) intended to humidify respired gases for tracheostomized patients with minimum tidal volumes of…
BS EN ISO 27427:2023
ДействующийAnaesthetic and respiratory equipment. Nebulizing systems and components.
IEC 60601-1-2:2014
ДействующийMedical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Co…
Overview IEC 60601-1-2:2014, published by the International Electrotechnical Commission (IEC), defines the requirements and tests relating to the basic safety and essential performance of medical ele…