ISO 14243-3:2004
Implants for surgery — Wear of total knee-joint prostheses — Part 3: Loading and displacement parameters for wear-testing machines with displacement control and corresponding environmental conditions for test
Implants for surgery — Wear of total knee-joint prostheses — Part 3: Loading and displacement parameters for wear-testing machines with displacement control and corresponding environmental conditions for test
- Статус документа:
- Отменён
- Формат:
- Электронный (PDF)
- Количество страниц:
- 15
- Дата публикации:
- 17 сентября 2004 г.
- Издание:
- ISO IS 14243 edition 1 version 1
- ICS:
- 11.040.40
ISO 14243-3:2004 specifies relative movement between articulating components, the pattern of the applied force, speed and duration of testing, sample configuration and test environment to be used for the wear testing of total knee-joint prostheses in wear-testing machines having axial load control, flexion/extension angular motion control, AP displacement control and tibial rotation control. The kinematics of ISO 14243-3:2004 may not be applicable to knee designs with a high degree of constraint, which could result in damage to the articulating components in the early stages of the test that would not be representative of clinical service.
Технические детали
- Технический комитет
- ISO/TC 150/SC 4 - Bone and joint replacements