ISO 21073:2019 - Microscopes - Confocal microscopes - Optical data of fluorescence confocal microscopes for biological imaging - defines standardized optical performance quantities for fluorescence confocal laser scanning microscopy (CLSM). The standard applies specifically to single-point confocal scanners using single-photon excitation and provides measurement and declaration rules so manufacturers and users can compare imaging performance reliably for biological specimens.
Key topics and technical requirements
- Resolution (lateral and axial): Defined as the full width at half maximum (FWHM) of the confocal point spread function (cPSF) measured at the centre of the field. Resolution measurements must use unprocessed data and a Gaussian fit.
- Strength of optical sectioning: Measured as FWHM of the response when scanning a thin fluorescent layer or reflective planar object through focus - a key metric for optical sectioning capability.
- Measurement conditions and test objects:
- Use point-like fluorescent microspheres smaller than half the expected resolution.
- Thin fluorescent layers for axial tests must be significantly thinner than the expected sectioning strength.
- Detection pinhole sizing specified in Airy units (AU); 1 AU is generally recommended.
- Sampling and field requirements:
- Pixel size should be at least ten times smaller than the theoretical resolution.
- Scanned field should be at least ten times larger than the theoretical resolution.
- Data integrity and signal path:
- Recordings must be linear (no detector/excitation saturation).
- Include full cPSF range (max and min) to avoid clipped FWHM estimates.
- Additional performance quantities covered in the standard (as listed in the document):
- Field uniformity and centring accuracy
- Co-registration accuracy (multi-channel alignment)
- Stability of illumination power
- Field number of the confocal scan optic
- Scanning frequency
- Declaration requirements: Measurements must be reported with test conditions (excitation/detection wavelengths, objective designation, pinhole size and reference wavelength, polarization, etc.).
Applications and who uses it
- Enables manufacturers to publish comparable optical data for confocal microscopes.
- Helps core facilities, imaging scientists, and life-science researchers to:
- Select instruments that meet required resolution and optical sectioning for fluorescence biological imaging.
- Monitor instrument performance and perform acceptance testing.
- Ensure reproducible quantitative imaging and multi-channel co-registration.
- Useful for procurement, method validation, and quality assurance in research labs, microscopy cores, and imaging-enabled diagnostics.