Ophthalmic instruments — Tonometers
Ophthalmic instruments — Tonometers
- Статус документа:
- Действующий
- Формат:
- Электронный (PDF)
- Количество страниц:
- 14
- Дата публикации:
- 29 сентября 2009 г.
- Издание:
- ISO IS 8612 edition 2 version 1
- ICS:
- 11.040.70
ISO 8612:2009, together with ISO 15004-1, specifies minimum requirements and the design compliance procedure for tonometers intended for routine clinical use in the estimation of intraocular pressure (IOP). ISO 8612:2009 takes precedence over ISO 15004-1, if differences exist.
Abstract
Overview
ISO 8612:2009 - Ophthalmic instruments - Tonometers defines minimum design and testing requirements for tonometers intended for routine clinical estimation of intraocular pressure (IOP). Applied together with ISO 15004-1, this International Standard establishes how manufacturers must demonstrate design compliance and verify individual instruments. Where differences exist, ISO 8612:2009 takes precedence over ISO 15004-1. Normative annexes provide the reference tonometer specifications (Annex A) and design compliance testing procedures (Annex B).
Key Topics and Requirements
- Scope and purpose: Specifies requirements for tonometers used to estimate IOP in clinical settings; recognizes that true IOP cannot be measured invasively and provides alternative reference methods.
- Design compliance testing: Manufacturers must show that test tonometer readings agree with a reference tonometer per defined tolerances. Acceptance is based on paired differences - not more than 5% of paired differences in each pressure range may exceed the tolerance.
- Performance tolerances and sample size:
- IOP ranges: 7–16 mmHg, >16–