IEC 61010-2-011:2019
Safety requirements for electrical equipment for measurement, control, and laboratory use - Part 2-011: Particular requirements for refrigerating equipment
Safety requirements for electrical equipment for measurement, control, and laboratory use - Part 2-011: Particular requirements for refrigerating equipment
- Статус документа:
- Действующий
- Формат:
- Электронный (PDF)
- Количество страниц:
- 155
- Дата публикации:
- 22 марта 2019 г.
- Издание:
- IEC IS 61010 edition 2 version 1
- ICS:
- 19.080
IEC 61010-2-011:2019 specifies particular safety requirements for the following types a) to c) of electrical equipment and their accessories, wherever they are intended to be used, whenever that equipment incorporates refrigerating systems as an integral part of, or separate from, the equipment and the equipment is in direct control of the refrigerating system. This document details all the requirements when up to 150 g of flammable refrigerant are used per stage of a refrigerating system. Additional requirements beyond the current scope of this document apply if a refrigerant charge of flammable refrigerant exceeds this amount. a) Electrical test and measurement equipment This is equipment which by electromagnetic means tests, measures, indicates or records one or more electrical or physical quantities, also non-measuring equipment such as signal generators, measurement standards, power supplies for laboratory use, transducers, transmitters, etc. b) Electrical industrial process-control equipment This is equipment which controls one or more output quantities to specific values, with each value determined by manual setting, by local or remote programming, or by one or more input variables. c) Electrical laboratory equipment This is equipment which measures, indicates, monitors, inspects or analyses materials, or is used to prepare materials, and includes in vitro diagnostic (IVD) equipment. This second edition cancels and replaces the first edition published in 2016. This edition constitutes a technical revision. This edition includes the following significant technical changes with respect to the previous edition: a) alignment with changes introduced by Amendment 1 of IEC 61010-1:2010; b) introduction of new defined terms or modified terms to align with Part 2-012 and other source documents. Editorial changes to use small capitals only for defined terms. Note the difference of defined term in 4.3.2.101 and abnormal operation in 11.7.104.3 and 11.7.104.5; c) clarifications for cooling tests in 4.4.2.10; d) changes pertaining to the accurate employment of the following terms: temperature, operating temperature, application temperature, controlled temperature, room ambient and ambient temperature; e) use of defined term to replace cooling system; f) move text of 4.4.2.101 to 4.3.2.101, since the purpose of as defined, is to simulate failure of the ambient conditions of 1.4.1 but not of the of the equipment; g) use of the term equipment to replace unit, apparatus, appliance, where applicable; h) in 5.1.2 dd) for high and low sides for each stage are required only under ; i) use of defined term to replace normal operation; j) use of defined term to replace user. It has the status of a group safety publication in accordance with IEC Guide 104.
Abstract
Overview
IEC 61010-2-011:2019 specifies particular safety requirements for refrigerating equipment that is part of electrical equipment for measurement, control and laboratory use. It applies where the equipment (electrical test & measurement, industrial process-control, or laboratory/IVD equipment) incorporates a refrigerating system and the equipment directly controls that system. The standard covers refrigerant charges of flammable refrigerant up to 150 g per stage; larger charges require additional requirements outside this Part 2‑011 scope. This is Edition 2.0 (2019) and is used together with the latest IEC 61010‑1.
Key topics and technical requirements
IEC 61010-2-011 supplements IEC 61010‑1 by adding refrigerating‑specific rules. Main topics include:
- Scope and definitions for refrigerating systems and controlled/ambient temperatures (alignment with Part 1 and Part 2‑012).
- Tests: cooling performance tests, leakage and rupture assessments, and motor‑compressor temperature/pressure tests.
- Protection requirements: protection against electric shock, mechanical hazards, spread of fire, and HAZARDS from fluids and solid foreign objects.
- Temperature and heat resistance: maximum temperature limits for motor‑compressors and components.
- Gases and explosion: safeguards for liberated gases, flammability parameters and explosion/implosion risks (annexes provide guidance).
- Components and piping: requirements and parameters for pressure vessels, tubing and piping (incl. referenced parameters in informative annexes).
- Marking and documentation: required labels, information for equipment containing flammable refrigerants, and user/operator instructions.
- Safety interlocks and risk assessment: provisions for interlocks and guidance to perform RISK assessment for refrigerating systems.
- Annexes: informative and normative annexes cover leakage/rupture, hazards, component/piping requirements, and marking for flammable refrigerants.
Significant technical updates in the 2019 edition include alignment with Amendment 1 to IEC 61010‑1, clarified terminology, refined cooling tests, and standardized use of defined terms.
Practical applications and who uses this standard
IEC 61010-2-011 is used by:
- Manufacturers of laboratory equipment, environmental chambers, medical/IVD devices and test instruments that include refrigerating systems.
- Product safety engineers and compliance teams preparing CE/CB/other conformity evidence.
- Test labs and certification bodies performing safety testing and assessments for refrigerating equipment.
- Design engineers specifying compressors, piping and safety interlocks for small-charge systems (≤150 g flammable refrigerant per stage).
It helps ensure safe design, testing, marking, and documentation to mitigate electrical, mechanical, thermal and flammability hazards.
Related standards
- IEC 61010‑1 (general safety requirements for measurement, control and laboratory equipment) - Part 1 is normative and must be used in conjunction with Part 2‑011.
- IEC 61010‑2‑010 and IEC 61010‑2‑012 (related Part 2 standards for heating and combined heating/refrigerating systems).
Keywords: IEC 61010-2-011, refrigerating equipment safety, flammable refrigerant, laboratory equipment safety, refrigerating system requirements, IEC standard.
Технические детали
- Технический комитет
- TC 66 - Safety of measuring, control and laboratory equipment
- SKU
- IEC 61010-2-011:2019
Похожие стандарты
Стандарты, упомянутые в описании
IEC 61010-1:2010
ДействующийSafety requirements for electrical equipment for measurement, control, and laboratory use - Part 1: General r…
Overview IEC 61010-1:2010 is an international standard published by the International Electrotechnical Commission (IEC) that sets general safety requirements for electrical equipment used in measurem…
IEC 61010-2-010:2019
ДействующийSafety requirements for electrical equipment for measurement, control and laboratory use - Part 2-010: Partic…
Overview - IEC 61010-2-010:2019 (Laboratory equipment for the heating of materials) IEC 61010-2-010:2019 is the Part 2 standard in the IEC 61010 series that specifies particular safety requirements f…
IEC 61010-2-012:2016
ДействующийSafety requirements for electrical equipment for measurement, control, and laboratory use - Part 2-012: Parti…
Overview IEC 61010-2-012:2016 is an international standard published by the International Electrotechnical Commission (IEC), focusing on safety requirements for electrical equipment used in climatic…