ISO 14243-5:2019
Implants for surgery — Wear of total knee prostheses — Part 5: Durability performance of the patellofemoral joint
Implants for surgery — Wear of total knee prostheses — Part 5: Durability performance of the patellofemoral joint
- Статус документа:
- Действующий
- Формат:
- Электронный (PDF)
- Количество страниц:
- 16
- Дата публикации:
- 23 мая 2019 г.
- Издание:
- ISO IS 14243 edition 1 version 1
- ICS:
- 11.040.40
This document specifies the relative angular movement between articulating patellofemoral joint components, the pattern of the applied force, speed and duration of testing, sample configuration and test environment to be used for the durability testing of total knee-joint prostheses in wear-testing machines with load control and displacement.
Abstract
Overview
ISO 14243-5:2019 - "Implants for surgery - Wear of total knee prostheses - Part 5: Durability performance of the patellofemoral joint" defines a standardized durability test method for the patellofemoral articulation of total knee prostheses. The standard specifies the relative angular movements, force patterns, speeds and duration, sample configuration and test environment for wear-testing machines operating under load control and displacement. It is focused on qualitative visual assessment of UHMWPE patella components articulating against femoral components to detect delamination, cracking or other failure modes.
Key topics and technical requirements
- Test principle: Cyclic waveforms combining a squatting sub-cycle and two low-flexion sub-cycles to simulate patellofemoral kinetics and kinematics until failure or the cycle limit.
- Cycle parameters: Squat sub-cycle up to 120° tibiofemoral flexion at 1/3 Hz with compressive load range 0.5–3× body weight; low-flexion sub-cycles 0–58° at 1.1 Hz ±0.1 Hz with loads 0.3–0.5× body weight. Total cycle time ≈ 4.82 s.
- Cycle limit: Tests may be stopped if no failure is observed by 50 000 cycles (updated from prior editions).
- Specimen selection: Use worst-case geometry (typically largest femoral with smallest patellar component). At least 5 polymeric patellar components recommended for qualitative assessment.
- Specimen preparation: Patella mounted with rigid fixture and fastened or cemented using PMMA (per ISO 5833) to avoid cushioning; optional artificial aging per ASTM F2003 when appropriate.
- Apparatus & setup: Requires wear-testing machines capable of producing specified angular/translational displacements and loads, CAD-derived kinematic waveforms, friction-sensitive mediolateral force fixtures and measurement; test fluid per ISO 14243‑1/3 with optional additives (e.g., sodium azide, EDTA).
- Reporting: Document specimen geometry justification (FEA or experimental), fixture/mounting method, cycle waveforms, environmental conditions and observed failure modes.
Applications and who uses it
ISO 14243-5:2019 is intended for:
- Orthopaedic implant manufacturers - design verification and durability validation of patellofemoral components.
- Biomechanical testing laboratories - standardized wear/durability testing and comparative evaluations.
- R&D and FEA engineers - defining worst-case geometries and deriving kinematic inputs.
- Regulatory and quality teams - preclinical evidence for submissions and risk assessment.
- Clinical researchers - correlating laboratory failure modes with retrieved implants.
Practical uses include pre-market durability testing, design comparisons, material selection (UHMWPE), and failure-mode investigations.
Related standards
- ISO 14243-1, -2, -3 (wear testing parameters, measurement methods and displacement control)
- ISO 7207-1 (knee component terminology and dimensions)
- ISO 5833 (acrylic resin cements / PMMA)
- ASTM F2003 (accelerated aging of UHMWPE)
Keywords: ISO 14243-5:2019, patellofemoral joint, total knee prostheses, durability testing, wear-testing machine, UHMWPE patella, load control, PMMA, ASTM F2003, worst-case geometry, 50 000 cycles.
Технические детали
- Технический комитет
- ISO/TC 150/SC 4 - Bone and joint replacements
- SKU
- ISO 14243-5:2019
Похожие стандарты
Стандарты, упомянутые в описании
ISO 5833:2002
ДействующийImplants for surgery — Acrylic resin cements
Overview ISO 5833:2002 - "Implants for surgery - Acrylic resin cements" sets international requirements for acrylic resin (bone) cements used to fix internal orthopaedic prostheses. The standard appl…
ISO 14243-1:2009
ДействующийImplants for surgery — Wear of total knee-joint prostheses — Part 1: Loading and displacement parameters for…
Overview ISO 14243-1:2009 - Implants for surgery - Wear of total knee-joint prostheses - Part 1 describes the loading, displacement and environmental parameters for laboratory wear testing of total k…
ISO 7207-1:2007
ДействующийImplants for surgery — Components for partial and total knee joint prostheses — Part 1: Classification, defin…
Overview ISO 7207-1:2007 - Implants for surgery - Components for partial and total knee joint prostheses - Part 1: Classification, definitions and designation of dimensions provides a standardized fr…