ISO 16063-17:2016
Methods for the calibration of vibration and shock transducers — Part 17: Primary calibration by centrifuge
Methods for the calibration of vibration and shock transducers — Part 17: Primary calibration by centrifuge
- Статус документа:
- Действующий
- Формат:
- Электронный (PDF)
- Количество страниц:
- 7
- Дата публикации:
- 30 мая 2016 г.
- Издание:
- ISO IS 16063 edition 1 version 1
- ICS:
- 17.160
ISO 16063 comprises a series of documents dealing with methods for the calibration of vibration and shock transducers. ISO 16063-17:2016 lays down detailed specifications for the instrumentation and procedure to be used for primary calibration of accelerometers using centrifuge calibration. ISO 16063-17:2016 is applicable to rectilinear accelerometers with zero-frequency response, mainly of the strain gauge or piezoresistive type, and to primary standard and working transducers. It is applicable for a calibration range from 10 m/s2 to 20 000 m/s2 (higher accelerations possible) at 0 Hz. The limits of uncertainty applicable are ±1 % of reading.
Abstract
Overview
ISO 16063-17:2016 - "Methods for the calibration of vibration and shock transducers - Part 17: Primary calibration by centrifuge" specifies instrumentation, environmental conditions and procedures for the primary calibration of accelerometers using centrifuge (rotational) methods. It applies to rectilinear accelerometers with zero‑frequency response (mainly strain‑gauge and piezoresistive types) including primary standard and working transducers. The standard covers a calibration range from 10 m/s² to 20 000 m/s² (higher accelerations possible) with applicable limits of uncertainty of ±1 % of reading.
Key topics and technical requirements
- Calibration principle: generation of known centripetal acceleration by rotating the accelerometer at controlled radii and rotational frequencies (acceleration ∝ radius × (rotational frequency)²).
- Two methods:
- Method 1 - uses a measured rotation radius (radius uncertainty < ±0.15 %) and measured rotational frequency to calculate reference acceleration and sensitivity.
- Method 2 - used when radius cannot be measured with required uncertainty; uses two radial positions and matched output frequencies; distance between positions measured with uncertainty ≤ ±0.5 %.
- Apparatus requirements:
- Balanced rotational table/arm rotating about a vertical axis; levelling within ±0.5° for lower ranges and ±2° for higher accelerations.
- Rotational frequency stability within ±0.05 %.
- Transducer axis alignment within ±0.5°.
- Heavy, vibration‑isolated base and low hum/noise when rotating.
- Instrumentation performance:
- Rotational frequency measurement contribution (expanded, k=2) ≤ 0.05 %.
- Voltage measurement contribution (expanded, k=2) ≤ 0.01 %.
- Preferred test points: six acceleration values chosen from 10, 20, 50, 100, 200, 500 m/s² (and multiples of ten); nominal reference acceleration 100 m/s² (second choice 50 m/s²).
- Uncertainty reporting: combined and expanded uncertainties calculated per Annex A (coverage factor k = 2) and reported with calibration results.
Applications and users
ISO 16063-17 is used by:
- National metrology institutes and accredited calibration laboratories performing primary accelerometer calibration.
- Transducer manufacturers validating sensor sensitivity and linearity.
- Aerospace, automotive, defense and industrial test laboratories requiring traceable, high‑acceleration calibrations.
- Organizations implementing vibration test programs that need accurately calibrated sensors for shock and vibration characterization.
Keywords: ISO 16063-17, centrifuge calibration, accelerometer calibration, vibration transducers, primary calibration, strain gauge accelerometers, piezoresistive accelerometers, rotational frequency, measurement uncertainty.
Related standards
Технические детали
- Технический комитет
- ISO/TC 108 - Mechanical vibration, shock and condition monitoring
- SKU
- ISO 16063-17:2016
Похожие стандарты
Стандарты, упомянутые в описании
BS ISO 16063-17:2016
ДействующийMethods for the calibration of vibration and shock transducers. Primary calibration by centrifuge.
ISO 16063-22:2005
ДействующийMethods for the calibration of vibration and shock transducers — Part 22: Shock calibration by comparison to…
Overview ISO 16063-22:2005 - Methods for the calibration of vibration and shock transducers - Part 22: Shock calibration by comparison to a reference transducer - defines instrumentation and procedur…