ISO 16063-32:2016
Methods for the calibration of vibration and shock transducers — Part 32: Resonance testing — Testing the frequency and the phase response of accelerometers by means of shock excitation
Methods for the calibration of vibration and shock transducers — Part 32: Resonance testing — Testing the frequency and the phase response of accelerometers by means of shock excitation
- Статус документа:
- Действующий
- Формат:
- Электронный (PDF)
- Количество страниц:
- 11
- Дата публикации:
- 24 августа 2016 г.
- Издание:
- ISO IS 16063 edition 1 version 1
- ICS:
- 17.160
ISO 16063-32:2016 lays down detailed specification for instruments and procedures of testing the frequency and the phase response of accelerometers by means of shock excitation. It applies to the accelerometers of the piezoelectric, piezoresistive and variable capacitance types with the damping ratio less than critical and in the frequency range up to 150 kHz. The method presumes that the frequency and the phase responses of the accelerometer under test gained by this method are the best possible characteristics for the mounted accelerometer on the condition that the recommendations for mechanical mounting of accelerometer stated in ISO 5348 are fulfilled and that the mass of the reference shock ball exceeds at least three times the mass of the accelerometer under test. Phase response of the accelerometer under test gained by this method is considered to be some "virtual" characteristic of accelerometer presuming that there is zero phase shift between the input and output signals at a frequency of 0 Hz. NOTE 1 It is intended that the user be aware that for the same accelerometer in the field application, the frequency and the phase responses might be different, depending on the mass and compliance of the test structure and the method of mounting. The method allows just a qualitative evaluation of the frequency and the phase response of accelerometers. NOTE 2 It is intended that the user does not try to get better resolution of the initial parts of the frequency and phase responses of the accelerometer under test than the dynamic range of the adequate characteristic provides it. The best use of the frequency and the phase responses of the accelerometer gained by this method are to get the best fit lines for the initial parts of the mentioned characteristics.
Abstract
Overview - ISO 16063-32:2016 (Resonance testing, accelerometer calibration)
ISO 16063-32:2016 specifies instruments and procedures for resonance testing of accelerometers by means of shock excitation. The standard applies to piezoelectric, piezoresistive and variable‑capacitance accelerometers with damping ratios less than critical, and covers frequency and phase response assessment up to 150 kHz. It provides practical requirements for setup, reference elements and data processing to obtain the best possible mounted accelerometer characteristics when mechanical mounting follows ISO 5348.
Key topics and technical requirements
- Scope and limits
- Frequency range: up to 150 kHz.
- Applicable transducer types: piezoelectric, piezoresistive, variable capacitance (damping < critical).
- Method provides a qualitative evaluation of frequency and phase response; field responses may differ due to mounting and structure mass.
- Reference shock ball
- Hardened and polished steel (≥ HRC50); flat threaded mounting surface.
- Mass of reference shock ball must be at least 3× the mass of the accelerometer under test.
- Typical ball diameters recommended (practical examples): D = 32 mm (for <100 kHz) and D = 19 mm (for <150 kHz).
- Measurement and uncertainty
- Recommended minimum frequency-domain resolution: 400 lines.
- Typical standard uncertainty for resonance frequency ≈ 0.5% (depends on time-record length).
- Damping ratio (typical piezoelectric ≈ 0.01) standard uncertainty ≈ 1% with ~80 dB dynamic range.
- Phase resolution example: ~5° with 40 dB phase noise suppression.
- Apparatus & environment
- Key items: impact/guide tube, signal analyser, conditioning amplifier, reference shock ball.
- Environmental guidance: room temperature 23 ± 5 °C, relative humidity < 90%.
- Data interpretation
- Phase response is treated as a “virtual” characteristic (zero phase shift assumed at 0 Hz).
- Best used to derive fit lines for initial parts of frequency/phase characteristics.
Applications and users
- Uses: resonance testing, frequency response and phase response checks during production, laboratory calibration, R&D and quality assurance of accelerometers and vibration sensors.
- Who benefits: calibration laboratories, test houses, manufacturers of accelerometers, metrology institutes, and engineers performing sensor qualification for high‑frequency or shock environments.
Related standards
- ISO 5348 - Mechanical mounting of accelerometers (referenced).
- Other parts of ISO 16063 series cover primary calibration methods and related test procedures (e.g., ISO 16063‑1, ‑21, ‑22, ‑31).
Keywords: ISO 16063-32, resonance testing, accelerometer calibration, frequency response, phase response, shock excitation, reference shock ball, piezoelectric accelerometers.
Технические детали
- Технический комитет
- ISO/TC 108 - Mechanical vibration, shock and condition monitoring
- SKU
- ISO 16063-32:2016
Похожие стандарты
Стандарты, упомянутые в описании
ISO 5348:1998
ОтменёнMechanical vibration and shock — Mechanical mounting of accelerometers
ISO 16063-22:2005
ДействующийMethods for the calibration of vibration and shock transducers — Part 22: Shock calibration by comparison to…
Overview ISO 16063-22:2005 - Methods for the calibration of vibration and shock transducers - Part 22: Shock calibration by comparison to a reference transducer - defines instrumentation and procedur…
ISO 16063-11:1999
ДействующийMethods for the calibration of vibration and shock transducers — Part 11: Primary vibration calibration by la…
Overview ISO 16063-11:1999 - "Methods for the calibration of vibration and shock transducers - Part 11: Primary vibration calibration by laser interferometry" defines primary calibration procedures f…