ISO 5832-2:2018
Implants for surgery — Metallic materials — Part 2: Unalloyed titanium
Implants for surgery — Metallic materials — Part 2: Unalloyed titanium
- Статус документа:
- Отменён
- Формат:
- Электронный (PDF)
- Количество страниц:
- 3
- Дата публикации:
- 21 марта 2018 г.
- Издание:
- ISO IS 5832 edition 4 version 1
- ICS:
- 11.040.40
ISO 5832-2:2018 specifies the characteristics of, and corresponding test methods for, unalloyed titanium for use in the manufacture of surgical implants. Six grades of titanium based on tensile strength are listed in Table 2. NOTE The mechanical properties of a sample obtained from a finished product made of this metal do not necessarily comply with those specified in ISO 5832-2:2018.
Abstract
Overview
ISO 5832-2:2018 - "Implants for surgery - Metallic materials - Part 2: Unalloyed titanium" specifies the required chemical, microstructural and mechanical characteristics, plus corresponding test methods, for unalloyed titanium intended for surgical implant manufacture. The fourth edition (2018) defines six titanium grades based on tensile strength, sets compositional limits for elements such as oxygen, nitrogen, hydrogen and iron, and clarifies test and sampling practices for conformity assessment.
Key topics and technical requirements
- Material grades: Six grades (including an ELI grade) distinguished by minimum tensile and yield strength and elongation - e.g., tensile minima range from about 200 MPa (ELI) up to 680 MPa (cold-worked Grade 4B) as specified in Table 2.
- Chemical composition: Maximum limits for impurities (oxygen, nitrogen, hydrogen, carbon, iron) are defined; hydrogen must be measured after final heat treatment and pickling.
- Microstructure: Annealed material must be uniform with grain size no coarser than No. 5; at ×100 magnification no inclusions or foreign phases shall be visible.
- Mechanical properties: Tensile, yield, elongation and reduction of area requirements are specified; sheet and strip have mandrel-bend requirements (no cracking on outside surface).
- Test methods: Recognised standards referenced include ISO 6892-1 (tensile testing), ISO 643 or ASTM E112 (grain size), ISO 7438 (bend test) and routine analytical procedures for composition.
- Conformity notes: The standard warns that mechanical properties of samples taken from finished products do not necessarily comply with the listed values unless processed and tested accordingly.
Applications and who uses this standard
ISO 5832-2 is essential for:
- Orthopaedic and dental implant manufacturers selecting and qualifying unalloyed titanium for screws, plates, dental fixtures and other load-bearing devices.
- Materials suppliers and stockholders certifying batch composition and mechanical performance.
- Testing laboratories performing tensile, bend and microstructural analyses to prove conformity.
- Biomedical engineers and design teams specifying appropriate titanium grades by mechanical and biological considerations.
- Regulatory bodies and procurement teams referencing standardized material specifications for device approvals and tenders.
This standard supports reliable implant manufacturing by ensuring material consistency, traceability and predictable mechanical behavior for unalloyed titanium used in surgical applications.
Related standards
- Other parts of the ISO 5832 series (implants for surgery - metallic materials)
- ISO 6892-1, ISO 643, ISO 7438, ASTM E112
- ASTM F67 (referenced for reduction-of-area guidance)
Keywords: ISO 5832-2, unalloyed titanium, surgical implants, titanium grades, mechanical properties, chemical composition, implant manufacturing, biocompatibility, test methods.
Технические детали
- Технический комитет
- ISO/TC 150/SC 1 - Materials
- SKU
- ISO 5832-2:2018
Похожие стандарты
Стандарты, упомянутые в описании
ISO 6892-1:2019
ДействующийMetallic materials — Tensile testing — Part 1: Method of test at room temperature
Overview ISO 6892-1:2019 - "Metallic materials - Tensile testing - Part 1: Method of test at room temperature" specifies a standardized method for performing tensile tests on metallic materials at ro…
ISO 6431:1992
ОтменёнPneumatic fluid power — Single rod cylinders, 1 000 kPa (10 bar) series, with detachable mountings, bores fro…
ISO 7438:2016
ОтменёнMetallic materials — Bend test
ISO 5832-2:2025
ДействующийImplants for surgery — Metallic materials — Part 2: Unalloyed titanium
Overview ISO 5832-2:2025 is an internationally recognized standard specifying the requirements for unalloyed titanium used in the manufacture of surgical implants. This fifth edition updates the prev…
ISO 5832-3:2021
ДействующийImplants for surgery — Metallic materials — Part 3: Wrought titanium 6-aluminium 4-vanadium alloy
Overview ISO 5832-3:2021 - "Implants for surgery - Metallic materials - Part 3: Wrought titanium 6‑aluminium 4‑vanadium alloy (Ti‑6Al‑4V)" - specifies the chemical, microstructural and mechanical cha…